The Development of Flexible Scintillation Panels Based on Chalcogenide and Oxide Phosphors for Advanced X-Ray Scanners and Tomographs
DOI:
https://doi.org/10.15407/scin12.06.039Keywords:
composite, detector, fine crystalline scintillator, ionizing radiation, radiographic scanner, scintillation panelAbstract
The technology of flexible panels and dispersed scintillation elements for X-rays registration with high uniformity of scintillation parameters (mean deviation no more than 2%) and low cost is developed. Parameters of flexible scintillation panels have been optimized, which enable their obtaining with high spatial resolution. The panels can be used as intensifying screens in medical and industrial radiography. Variants of the performance of the dual energy X-ray detectors, working effectively in X-ray energy region (from 20 to 100 keV) are proposed. Possible areas of their application - multienergy scanners and medical computer tomography.
References
Riudiger Y.G. O primenenii siney i zelenoy system vi zualizacii izobrasheniia v rentgenographii. Medicinskaja technika. 2004, 6: 40-45 [in Russian].
Gurvich A.M. Rentgenoluminophory i rentgenovskie ekrany. Moskva: Atomizdat, 1976. 63 s. [in Russian].
Carel W E van Eijk. Inorganic scintillators in medical ima ging. Physics in medicine and biology. 2002, 47: 85-88. https://doi.org/10.1088/0031-9155/47/8/201
Ryzhikov V.D. Scintilliazionnye kristally polu pro vod ni kovych soedinenii АIIBIV. Poluchenie, svoistva, primenenie. Moskva: NIITECHIM, 1989. 127 s. [in Russian].
Atroshenko L.V., Burachas S.F., Halchynetskyi L.P., Grinev B.V., Ryzhikov V.D., Starzhynskyi N.H. Kristally scintillatorov i detectory ionoziruiushchich izlucheniy na ich osnove. Kyiv: Naukova dumka, 1998. 310 s. [in Russian].
Derenzo S.E., Bourret-Courchesne E., Weber M.J., Klintenberg M.K. Scintillation studies of CdS(In): effects of various semiconductor doping strategies. Nuclear Instruments and Methods in Physics Research. 2005. V. 537. P. 261-265. https://doi.org/10.1016/j.nima.2004.08.022
Nikl M. Scintillation detectors for X-rays. Measurements science and technology. 2006 17: R37 - R54.
https://doi.org/10.1088/0957-0233/17/4/R01
Moses W.W. Scintillation requirements for medical imaging. Proceedings of the Fifth International Conference on Inorganic Scintillators and their Applications, Moscow State University, 2000, pp. 11-21.